PR-655/670 User Manual
120
Setup
Command
Description
SE
Select Exposure Time
Enter the Exposure (Integration) time for the next measurement in milliseconds. Possible
values are 6
–
6,000 (6 sec.) for
Standard Mode, and
6
-
30,000 (30 sec.) for
Extended
Mode.
See the H specifier for more information on setting
Standard
or
Extended
Modes.
To set the instrument to
Adaptive Exposure,
send
SE0
(ttttt = 0)
Syntax
: SEttttt[CR]
Where
: ttttt = exposure time in milliseconds
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Standard and Extended modes apply only to PR-670. PR-655 exposure range is 3 to
6,000 ms
SF
Aperture Select (PR-670 only)
Select the aperture to be used for the next measurement.
Syntax
: SFa[CR]
Where
: a = aperture code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: See Data Code 117 for details on aperture codes.
SG
Speed Mode (PR-670 only)
Select the Speed Mode for the next measurement. Choices are Normal, 1X Fast, 2X Fast
and 4X Fast.
Syntax
: SGg[CR]
Where
: g = Gain
0 = Normal (DEFAULT),
1 = 1X for AC sources,
2 = 10X
3 = 100X
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...