PR-655/670 User Manual
113
Step 9 -
The next screen is a form used for entering RS232 communications. Since the USB driver sets
these parameters, no changes are required. Click on
OK.
FIGURE 124 - WINDOWS XP COM PROPERTIES WINDOW
Step 10 -
The main HyperTerminal appears. Type
PHOTO
(case sensitive).
Step 11 -
REMOTE MODE
is displayed in the HyperTerminal window. The
PR-655/670 is now ready to accept Remote Control Commands.
Step 12 -
Type
E
then press
Enter
to put the PR-655/670 in
Echo
or
Full Duplex
mode This enables display
of characters sent to the instrument on the HyperTerminal window.
E
NTERING
R
EMOTE
M
ODE
When communicating with the PR-655/670 using remote mode commands in the following environments, C++,
Visual Basic, MatLab, and LabView please note that
single characters and not Strings
must be sent to the
instrument. Some examples are given below.
Entering Remote Mode
To enter remote mode “P”
“H”
“O”
“T”
“O” must be sent as single characters and not as a single string.
Making a Measurement