PR-655/670 User Manual
124
M
EASUREMENT AND
D
ATA
S
END
C
ODES
Measurement and Data Send Codes are used to measure (
M
Command) and then specify returned data or
acquire values without making a measurement (
D
command).
Either a
D
or an
M
can precede any of the following codes. For example,
M1 or D1
–
M602 or D602
. If an M
command is sent, a measurement will always be made even if the Data Code does not request measured values.
For example, if
M116
is sent to the instrument, a measurement is made then a list of accessories is returned
from the instrument.
In the following table,
qqqqq
is the returned error code. If
qqqqq
is all zeros (00000) no error has occurred
during the request. All other values for
qqqqq
relate to an error condition. Refer to the Remote Mode Error
Code section of the manual for a complete list of error codes and their meanings.
UUUU
in the output format is
the photometric unit type of the measurement per the following table:
Type
Code
Units
Luminance
111
fL
cd/m²
Radiance
11
W/sr/m²
Illuminance
112
fc
lux
Irradiance
12
W/m²
Luminous Intensity
113
mcd
Radiant Intensity
13
W/sr
Luminous Flux
114
lumens
Radiant Flux
14
Watts
TABLE 9 - PHOTOMETRIC UNITS CODES
Note:
Both qqqqq and UUUU are contained in every output and are not annotated below.
All data fields are fixed length (except where otherwise noted) and comma delimited. Commas also serve as
placeholders for empty fields.
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...