Setup Commands are used to specify instrument and measurement parameters for the next measurement. To
specify more than one parameter, sequential setup commands may be sent to the instrument before the
measurement is initiated.
The default parameters are those used during the measurement immediately before Remote Mode operations
are initiated. If Remote Mode operations are under way, the default setup values for the upcoming measuring
are those defined in the last Setup (S_ command).
Setup
Command
Description
SA
Select Add-on Accessory 1
An Add-on accessory is one that is used in conjunction with a primary accessory. For
example, a neutral density filter (Add-on Accessory) used with the MS-75 (Primary
Accessory). Up to 3 Add-on accessories can be specified for a measurement.
Syntax
: SAn[CR]
Where
: n = Accessory code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.
SB
Select Add-on Accessory 2
An Add-on accessory is one that is used in conjunction with a primary accessory. For
example, a neutral density filter (Add-on Accessory) used with the MS-75 (Primary
Accessory). Up to 3 Add-on accessories can be specified for a measurement.
Syntax
: SBn[CR]
Where
: n = Accessory code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...