PR-655/670 User Manual
58
H
OW TO
M
AKE
L*
U
*
V
*
OR
L*
A
*
B
*
M
EASUREMENTS
1.
Touch the
▼
or
▲
icons in the
Measurement Mode
field to select
L*u*v*
or
L*a*b*
.
2.
To make
L*u*v*
or
L*a*b*
measurements, it is necessary to first select a white reference (
Illuminant
).
Touch
Select a
t th
e bott
om of the Illuminant column. A screen similar to the following ap
pears:
FIGURE 50 - L*U*V* SETUP SCREEN
FIGURE 51 - FILE RECALL SCREEN
To access factory stored illuminants, touch the
▼
or
▲
icons in the current view field until standard illuminants appears.
An illuminant
may also be a previously measured and stored file
–
for example a florescent lamp. If it is desirable to use a
previous measurement then select internal ram or external SD card (if an SD card is inserted).
OR
Press the
MEASURE
button then touch
Back
then
Recall
to accept the measured data.
FIGURE 52 - STANDARD ILLUMINANTS SCREEN
3.
If using a pre-stored illuminant, at the top of the screen that pops up, the title of the
lluminant appears.
Touch
Next
(or
Previous
) to scroll through available choices.
4.
Touch
Back.
5.
Touch
Recall
to select the illuminant of choice. The
Standard Illuminant
screen updates to show the
data of the illuminant selected.
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...