107
The host will attempt to re-connect with the PR-655/670. When connected the above icon will change to green
and a Connected indicator will be shown. The instrument will also indicate a successful link by displaying a
green Bluetooth icon
and the name of the Remote Host (in Hexadecimal).
RS-232
(
OPTIONAL
)
The traditional RS-232 I/F allows for interfacing to ATE environments and older generation PC systems. The
below diagrams depicts the connection between the PR-655/670 and the host computer. To establish a
connection between the PR-655/670 simply open the appropriate port (using the specified protocol settings)
and enter “
PHOTO
”, no other hardware handshakes are necessary.
FIGURE 116 - PR-655/670 RS-232 OPTION
H
ARDWARE
P
ROTOCOL
The RS-232 hardware protocol settings are:
•
Baud Rate
: Selectable (9600, 19.2 Kbps 38.4 Kbps, 57.6 Kbps, 115.2
1
Kbps)
•
Parity
:
None
•
Data
Bits
:
8
•
Stop
Bits
: 1
1
Default baud rate unless in PR-650 emulation mode.