PR-655/670 User Manual
66
1.
Touch the
▲▼
icons adjacent to
Match Method
to display either
Y
x,y
or
x,y.
2.
Enter the desired values for
Y, x
and
y
in the fields directly below
Reference Data
by touching any of the
fields. This action brings up a data entry screen like the following example for
Y
:
FIGURE 60 -
R
G
B
DATA ENTRY
Note: If
xy
is selected, data entry for
Y
is unavailable.
3.
Enter the value for
Y,
and then touch
Back.
4.
Repeat for
x
and
y.
M
AKING AN
R
G
B
M
EASUREMENT
When all parameters have been successfully set, white point calibration can commence.
1.
Press the
MEASURE
button to begin the
R
G
B
measurement sequence. A screen similar to the following
appears: