PR-655/670 User Manual
34
To set the
SYNC
mode:
1.
Make sure
Adaptive
is set to
Enabled.
2.
Touch the
▼
or
▲
icon in the
SYNC
field to toggle between
None, Manual,
and
Automatic.
3.
If
Manual
is selected, touch the field adjacent to
Frequency to
access the following data entry screen
:
FIGURE 34 - SYNC PERIOD DATA ENTRY SCREEN
4.
Touch the numeric keys to enter the SYNC period. The range is 20 to 400 Hz.
5.
Touch
Back
when finished.
S
ETTING
S
MART
D
ARK
M
ODE
(PR-670
O
NLY
)
During a measurement the PR-655 and PR-670 make two separate measurements
–
one of the optical signal
(
Light
), and a second measurement of the detector dark current (
Dark
). By enabling the
Smart Dark
feature, the
PR-670 will attempt to use the same dark current values for more than one measurement thus reducing total
measurement time. The
Smart Dark
feature is used when the instrument determines that the (second or
successive) measurement in a series is using the same exposure time as the first measurement. If a fixed
exposure time is used for a measurement,
Smart Dark
will be used until the exposure time is changed.
Smart Dark
is a useful tool for applications such as display uniformity, or other testing where multiple
measurements of the same light level are performed.
To enable
Smart Dark:
1.
Touch the field next to
Smart dark
to toggle between
Enabled
and
Disabled.
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...