6 | Installation
INFICON
58 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
CAUTION
Do not clean copper or stainless steel tubing with solvent.
The solvent can enter Micro GC Fusion and cause damage to the instrument.
CAUTION
Do not use Teflon® or Tygon® tubing for carrier gas.
• Assortment of male and female Swagelok fittings for regulator, tubing, etc.,
including nuts and front and back ferrules. Typical size is 1/8 in. but adaptors for
1/4 in. to 1/8 in. may be needed if 1/4 in. tubing is used
• 1/8 in. tee fitting and 1/8 in. ball valve, depending on connection requirements
6.1.3 Calibration Gas Requirement
• Commercially supplied calibration gas specific to the analysis
Calibration gas must be available at the time of installation.
Calibration gas must closely resemble the composition of the actual process
gas.
• A regulator, if necessary, to reduce calibration gas pressure below 69 kPa (10 psi)
at the Micro GC Fusion inlets. The regulator must be clean, grease-free and non-
venting
• Copper or stainless steel tubing (pre-cleaned) 1/16 in. OD for calibration and
sample gas connections between the tank regulator output and Micro GC Fusion
sample inlet
Checkout gas mixtures are only used for instrument installation and startup.
Checkout gas mixtures supplied by INFICON are not calibration gases.
When the instrument is configured with a 1/16 in. Swagelok front sample inlet,
the ship kit includes a 61 cm (24 in.) 1/16 in. OD piece of stainless steel
sample tubing and nuts and ferrules for both ends.
Содержание Micro GC Fusion 2-Module
Страница 1: ...Operating Manual Micro GC Fusion Gas Analyzer 074 594 P1H ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 13: ...INFICON Declaration of Conformity 2 074 594 P1H Micro GC Fusion Operating Manual 13 319 ...
Страница 318: ......
Страница 319: ...074 594 P1H 2019 ...