7 | Connecting to a Sample
INFICON
148 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
7.9 Sampling Using the Integrated Sample
Conditioner
The optional Integrated Sample Conditioner is factory installed. It maintains the
incoming sample at 100°C (212°F) and reduces the sample pressure from up to 6,895
kPa (1,000 psi) down to 83 kPa (12 psi) prior to the sample being injected into the GC
module.
WARNING
Gas pressure on the HIGH PRESSURE PURGE VENT may be up to 448 kPa
(65 psi) . Extreme caution must be taken to vent the high pressure sample to
avoid personal injury.