INFICON
Product Overview | 5
074-594-P1H Micro GC Fusion Operating Manual
45 / 319
5.10.2 Sample Inlet
Sample Inlet
Specification
Front inlet
Heated, 30-140°C (86-284°F), 1/16 in.
Swagelok
Maximum sample pressure
172 kPa (25 psi)
Recommended sample pressure
Ambient to 69 kPa (10 psi)
5.10.3 Integrated Sample Conditioner
Integrated Sample Conditioner
Specification
Front inlet
Heated, 100°C (212°F), 1/8 in. quick
connect
Maximum sample pressure
6,895 kPa (1,000 psi)
Minimum sample pressure
34.5 kPa (5 psi) suggested for optimal
purging
5.10.4 Gas Liquid Separator
Gas Liquid Separator
Specification
Sample inlet
1/16 in. Swagelok
Maximum sample pressure
172 kPa (25 psi)
5.10.5 Pressure Reducer and Gas Liquid Separator
Pressure Reducer and Gas Liquid Separator
Specification
Sample inlet
1/8 in. Swagelok
Maximum sample pressure
3,450 kPa (500 psi)
5.10.6 Sample Injector
Sample Injector
Specification
Variable volume
Heated, 1 to 10 uL
Variable large volume
Heated, 1 to 30 uL
Fixed volume
Heated, 1.6 uL
Backflush to vent
Heated, 0.4 or 1 uL fixed, 7.1 uL maximum variable