6 | Installation
INFICON
76 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
6.6.3 List of Required Tools
The following tools are user supplied:
• 2 mm hex driver
• 5/16 in. wrenches
• 1/4 in. wrench
6.6.4 Installing the Gas Liquid Separator
The gas-liquid separator can be installed on the 2-module or the 4-module Micro GC
Fusion. The gas-liquid separator is factory pre-assembled, and can be easily mounted
onto Micro GC Fusion during instrument installation.
CAUTION
The gas-liquid separator filters out liquid droplets in a gas sample. It is not
intended for liquid injections.
WARNING
Gas pressure on the gas-liquid separator sample vent may be up to 172 kPa
(25 psi). Extreme caution must be taken when venting the high pressure
sample to avoid personal injury.
6.6.4.1 2-Module Chassis
To install the gas-liquid separator onto a 2-module Micro GC Fusion:
1 Press the On/Standby button once to shutdown Micro GC Fusion. This shutdown
procedure cools down the analytical columns before powering down and may
take 1 to 2 minutes.
2 Remove all electrical cables, including the power cable and the Ethernet cable.
3 Disconnect the sample gas connection(s), carrier gas connections and venting
connections.
4 Turn Micro GC Fusion onto its side.