Installation | 6
074-594-P1H Micro GC Fusion Operating Manual
129 / 319
• 2.0 hex driver
User supplied parts:
• vent tubing and fittings, if used
• 1/8 in. Swagelok type sample input tubing and female fittings
The cap plugs on the sample inlet and the sample vent of the pressure
reducer and gas-liquid separator are not shown in the figure. They must be
removed when performing this installation.
1 Turn off the instrument by following the system shutdown procedure in Shutdown
251].
2 Disconnect all the electrical cables, including the power cable and Ethernet
cable, from the instrument.
3 Disconnect the sample gas connection(s), the carrier gas connection(s) and the
venting connections from the instrument.
4 Ensure that Bracket C is positioned with two small holes at the bottom when in
the upright position. If not, follow the procedure in Change Mounting Bracket
Orientation [
137] to adjust the mounting bracket orientation.
Bracket C Bracket D