10 Insert the sample tubing with the Valco fittings into the fitting detail on the Genie
filter. Push the tubing all the way into the detail, so that it seats firmly. Manually
turn the nut into the detail until it is finger tight. Using the 1/4 in. open-ended
wrench, turn the nut 1/4 turn (90°) past the point where the ferrule first starts to
grab the tubing. The amount of force required can vary considerably due to the
friction between the nut and the treads and the composition and wall thickness of
the tubing used.
11 Put the 5/16 in. Swagelok nut (PN 059-0696) and the 1/16 in. ferrule set (PN
059-0697) on the other end of the sample tube.