10 | Routine Operation
INFICON
278 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
CAUTION
Make sure that the column module is inserted using the guiding pins and that
the connector is not damaged during installation.
6 Ensure the 100 pin connectors on the GC module controller board and the GC
module interface board are securely connected.
7 From the left side of the chassis frame, use a Phillips head screwdriver to tighten
the two captive screws to secure the GC module in place.
CAUTION
Do not over tighten the screws to avoid module damage.
8 Connect the sample line to the GC module injector manifold. Finger tighten the
connection and then use a 5/16 in. open end wrench to tighten it 1/4-turn past
finger tight.
Содержание Micro GC Fusion 2-Module
Страница 1: ...Operating Manual Micro GC Fusion Gas Analyzer 074 594 P1H ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 13: ...INFICON Declaration of Conformity 2 074 594 P1H Micro GC Fusion Operating Manual 13 319 ...
Страница 318: ......
Страница 319: ...074 594 P1H 2019 ...