INFICON
Checkout Gas Chromatograms | 14
074-594-P1H Micro GC Fusion Operating Manual
313 / 319
14.2.2 Universal Checkout Gas Using Argon Carrier Gas
Argon carrier gas is used for the analysis of hydrogen in a sample. Analysis of other
fixed gases is possible using argon carrier gas but with reduced sensitivity when
compared with helium carrier gas.
Peak Number
Compound
Concentration (mole %)
1
hydrogen
10.00
2
oxygen
1.00
3
nitrogen
81.24
4
methane
5.00
5
carbon monoxide
1.00
1
2
3
4
5
• Column Temperature Profile: 100°C (90 s hold)
• Column Pressure: 35 psi
• Inject (ms) Time: 30 ms
• Backflush Time: 25 s