9 | Software Operation
INFICON
230 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
10 Type 2 for a level 2 calibration.
11 Repeat steps 7 through 10 until you have reached the desired number of levels.
12 Click to save the sequence.
13 Run the Auto-Calibration sequence after all the calibration gases are connected.