3 Connect the 1/16 in. sample line to the 1 IN filter on the Valco stream selector.
4 While holding the filter with a 7/16 in. wrench, use a 1/4 in. wrench to turn the
union fitting. Over-tighten by 1/2 turn to swage the union fitting and ferrule onto
the 1/16 in. sample line.
5 Loosen the union fitting and then tighten it again.
6 Repeat steps 1-5 for sample lines to 2 IN filter, 3 IN filter, etc. until the desired
number of sample lines are connected.
7 Place a union fitting and ferrule onto a 1/16 in. OD stainless steel sample line.
8 Connect the 1/16 in. OD stainless steel tubing to the OUT TO GC union.
9 While holding the OUT TO GC fitting with a 1/4 in. wrench, use a 1/4 in. wrench
to turn the union fitting. Over-tighten by 1/2 turn to swage the union fitting and
ferrule onto the 1/16 in. sample line.
10 Loosen the union fitting and then tighten it again.