INFICON
Installation | 6
074-594-P1H Micro GC Fusion Operating Manual
95 / 319
6.12 The AUX I/O Connector
For some applications, an external triggering device may be required for sampling.
This can be accomplished by connecting an external triggering device to the Micro GC
Fusion AUX I/O port via a Remote Start Cable (PN G2801-60618). This cable has a
15 pin connector at one end. The other end can be fitted with a desired connector
type.
2-Module chassis:
4-Module chassis:
The AUX I/O connector pin outputs are listed in the table below.
Pin
Function
Color
1
Provides 5 mA remote input
Black
2
REMOTE_START input
White
3
REMOTE_START input
Red
4
Variable GND
Green
5
Provides "5V pull up" for logic command
Orange
6
Provides "5V pull up" for logic command
Blue
7
Contact closure output FAULT_OUT*
White/black
8
Contact closure output READY_OUT
Red/black
9
Provides 5 mA for remote input
Green/black
10
REMOTE_CANCEL input
Orange/black
11
REMOTE_CANCEL input
Blue/black