13 | Checkout Gases and Replacement Parts
INFICON
304 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
13.1.3 Refinery Gas Checkout Gas
Refinery gas checkout gas contains fixed gases, C1-C7 saturated hydrocarbons and
C2-C5 olefins in a nitrogen balance. See the table below for the component
concentrations.
Refinery gas checkout gas is not intended to be a substitute for a calibration
standard for any specific applications. INFICON recommends acquiring a
calibration standard gas that meets composition requirements appropriate for
the specific application.
Components
Concentration (Mole %)
hydrogen
10.00%
oxygen
0.19%
nitrogen
balance
carbon monoxide
1.00%
carbon dioxide
3.00%
methane
5.00%
ethane
4.00%
ethylene
2.00%
acetylene
1.00%
propane
2.00%
propylene
1.00%
isobutane
0.30%
n-butane
0.15%
1-butene
0.30%
isobutylene
0.15%
trans-2-butene
0.15%
cis-2-butene
0.30%
1,3-butadiene
0.30%
isopentane
0.10%
n-pentane
0.05%
cis-2-pentene
0.05%
trans-2-pentene
0.05%
methyl-2-butene
0.05%
n-hexane
0.05%
n-heptane
0.05%