RMV Operations Manual Revision Copyright illunis LLC, 2014
Page 55
4.2.1: Exposure:Trigger Modes
TPE: Triggered Programmed Exposure
TPE
mode uses the trigger pulse to start a pro-
grammed expose/readout cycle. The exposure is set by
the Transfer Pulse Delay (TPD) register. The TPD is set
in increments of 4,16,64 or 1024 pixel clock periods.
Serial Commands
Target
Index
Command
R/W
Description
04
03
Readout Mode Select
W
0x0001 = Trigger Program Exposure
02
04
Transfer Pulse Delay
R/W
0x0007 to 0xFFFF
04
1B
Transfer Pulse Width
R
0x0004 = TPW (Preset at factory)
02
00
Set Trigger ms
W
Set TR time in milliseconds * 100
02
01
Set Trigger us
W
Set TR time in us
02
00
Get Trigger ms
R
Return actual time in milliseconds * 100
02
01
Get Trigger us
R
Return actual time in us (0xFFFF = to large).
02
0E
TPD resolution
R/W
0x0000 = 4 clock periods
0x0001 = 16 clock periods
0x0002 = 64 clock periods
0x0003 = 1024 clock periods
Quick FAQ’s:
►Use the TPE mode to control exposure within the
RMV camera.
►The rising edge of the trigger pulse determines the
beginning of exposure.
►Multiple Cameras with the same trigger can be
slaved together for very exacting applications
►
The strobe signal can be found on the RMV power
connector and is a 3.3V LVTTL signal.
Trigger
Transfer Pulse
Strobe
SUB (Erase)
TST
TPW
TPD
TTSC
TEXP
Exposure
Read Out