![background image](http://html1.mh-extra.com/html/illunis/rmv-11002/rmv-11002_operation-manual_2044107036.webp)
RMV Operations Manual Revision Copyright illunis LLC, 2014
Page 36
System
status can be read from mode registers
and from the system built in test status register.
Serial Commands
Target
Index
Command
R/W Description
04
1b
System Registers
R
0x0000 = Read Pixels/line
0x0001 = Read Active pixels/line (in LVAL)
0x0002 = Read Lines per frame
0x0003 = Read Active lines per frame (in FVAL)
0x0004 = Read TPW
0x0005 = TRO Left Start
0x0006 = TRO Right Start
0x0007 = TRO Size
0x0008 = LVAL Start
0x0009 = LVAL Stop
0x000a = FVAL Start
0x000b = FVAL Stop
0x000c = CCD Type
0x000d = FPGA Revision
0x000e = Read TPD
0x000f = SNR Left
0x0010 = SNR Right
0x0011 = Crack detector position
0x0012 = Read Exposure value low
0x0013 = Read Exposure value hi
0x0014 = Read CRC
07
00
Camera
Parameters
R
0x0000 = Camera Model
0x0001 = Camera Hardware rev
0x0002 = Camera Serial Number
0x0003 = Micro firmware rev
0x0004 = FPGA major revision
0x0005 = Sensor Serial Number
0x0006 = Clock Rate
0x0007 = FPGA Sub/minor revision
0x0008 = Micro Sub/minor revision
Quick FAQ’s:
►These commands are very useful for determining
the state of the camera.
►The FPGA major and minor revision should be
checked by application software to match with ex-
pected levels.
►The clock rate must be divided by 100
3.4: Software ICD
System & Status