u
The first digit represents the surface:
u
1 is for top
u
2 is for bottom
u
The second digit represents the swath:
u
1 is for the first swath
u
2 is for the second swath
u
3 is for the third swath
u
The last 2 digits represent the tile, 01 through 16. Tile numbering starts with 01 at the output end of the
flow cell through 16 at the input end.
Figure 21
Tile Numbering
This example indicates a tile from the top surface of the flow cell, the second swath, and the seventh tile.
Thumbnail Images
You can configure the control software to generate thumbnail images in *.jpg file format. Thumbnail images
are generated for each cycle and base.
The control software collects images from 9 sections of a tile. The 9 images are combined into 1 thumbnail
image and can be used to troubleshoot a run. Thumbnail images are not suitable for image analysis, but can
be used for troubleshooting.
Figure 22
Thumbnail Image
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
63