Position
Reagent
Description
4
SBS Buffer 1 (SB1)
High Salt Buffer
5
SBS Buffer 2 (SB2)
Incorporation Wash Buffer
6
SBS Buffer 2 (SB2)
Incorporation Wash Buffer
7
CRM
Cleavage Reagent Mix
8
SBS Buffer 3 (SB3)
Cleavage Buffer
7
Put on a new pair of powder-free latex gloves.
8
Slide the reagent rack into the reagent compartment. Align the rack with the raised guide on the floor of
the compartment.
9
Lower the sippers into the SBS reagent bottles as follows.
a
Pull the sipper handle towards you and then lower it.
b
Inspect the sippers to make sure that they do not bend as they lower into the funnel caps.
c
Release the handle into the slot on the bottom end of the groove.
10 Select the PW1 (25 ml) loaded checkbox.
Load Indexing Reagents
1
Make sure that the paired-end rack is not in use on the adjacent flow cell for Read 2 resynthesis, Index 1
(i7) Read preparation, and Index 2 (i5) Read preparation.
2
Raise the sippers for the paired-end reagent rack as follows.
a
Pull the handle towards you and raise it.
b
Release the handle into the slot on the top end of the groove. Make sure that the handle rests
securely in the slot.
3
Slide the reagent rack out of the reagent compartment using the rack handle.
4
Remove the caps from each reagent tube and place the tube onto the rack in the associated numbered
position or matching label color.
Position
Reagent
Description
15
FDR
Fast Denaturation Reagent (contains formamide)
16
HP9*
Index Sequencing Primer i5
17
HP12
Index Sequencing Primer i7
Table 1 Single-Read Flow Cells
* HP9 is required for dual-indexed runs only. If HP9 is not used, load a 15 ml conical tube with 10 ml laboratory-grade water in position 16.
Position
Reagent
Description
10
FRM*
Fast Resynthesis Mix
15
FDR
Fast Denaturation Reagent (contains formamide)
17
HP12
Index Sequencing Primer i7
Table 2 Paired-End Flow Cells
* Load FRM in position 10 for dual-indexed runs on a paired-end flow cell. FRM is required in position 10 for all paired-end runs regardless of indexing
option.
5
If you are performing a paired-end run, skip steps
6
Place 15 ml conical tubes filled with 10 ml laboratory-grade water in unused positions on the paired-end
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
17
HiSeq 2000 System Guide