Combine the following volumes of HP3 and PW1:
u
For single-indexed runs—Transfer 3325 µl PW1 to an empty 15 ml conical tube, and then add 175 µl
HP3.
u
For dual-indexed runs—Transfer 3800 µl PW1 to an empty 15 ml conical tube, and then add 200 µl
HP3.
3
Invert to mix.
4
Centrifuge at 1000 rpm for 1 minute.
5
Set aside at room temperature.
Prepare HP8 or HP12
1
Invert to mix.
2
Centrifuge at 1000 rpm for 1 minute.
3
Set aside at room temperature.
Prepare RMR
1
Invert to mix.
2
Centrifuge at 1000 rpm for 1 minute.
Do not vortex.
3
Set aside on ice.
Prepare HP9
1
Invert to mix.
2
Briefly pulse centrifuge to collect droplets.
3
Set aside at room temperature.
Enter Run Parameters
Begin run setup by entering run parameters from a series of screens on the Run Configuration tab. The
software guides you through each screen to specify BaseSpace Sequence Hub connectivity, enter
consumable IDs, select indexing options, and record other parameters.
Integration Screen
The Integration screen provides the option to connect the run to BaseSpace Sequence Hub.
1
From the Welcome screen, select Sequence | New Run to open the Integration screen.
2
[Optional]
Connect to BaseSpace Sequence Hub as follows.
a
Select BaseSpace.
b
Select from the following BaseSpace options:
u
Storage and Analysis—Sends run data to BaseSpace Sequence Hub for remote monitoring and
data analysis. A sample sheet is required with this option.
u
Run Monitoring Only—Sends only InterOp files to BaseSpace Sequence Hub, which allows
remote monitoring of the run.
c
Log on to BaseSpace Sequence Hub using your MyIllumina account email and password.
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
30
HiSeq 2000 System Guide