7
Lower the sippers into the SBS reagent bottles as follows.
a
Pull the sipper handle towards you and then lower it.
b
Inspect the sippers to make sure that they do not bend as they lower into the funnel caps.
c
Release the handle into the slot on the bottom end of the groove.
8
Select the PW1 (25 ml) loaded checkbox.
Load Indexing Reagents
1
Make sure that the paired-end rack is not in use on the adjacent flow cell for Read 2 resynthesis, Index 1
(i7) Read preparation, and Index 2 (i5) Read preparation.
2
Raise the sippers for the paired-end reagent rack as follows.
a
Pull the handle towards you and raise it.
b
Release the handle into the slot on the top end of the groove. Make sure that the handle rests
securely in the slot.
3
Slide the reagent rack out of the reagent compartment using the rack handle.
4
Remove the caps from each reagent tube and place the tube onto the rack in the associated numbered
position or matching label color.
Position
Reagent
Description
17
HP8 or HP12
Index 1 (i7) Sequencing Primer Mix
18
HP3
Denaturation Solution
19
HT2
Wash Buffer
Table 3 Single-Indexed Run
Position
Reagent
Description
16
HP9
Index 2 (i5) SR Sequencing Primer Mix
17
HP8 or HP12
Index 1 (i7) Sequencing Primer Mix
18
HP3
Denaturation Solution
19
HT2
Wash Buffer
Table 4 Dual-Indexed Run on a Single-Read Flow Cell
Position
Reagent
Description
10
RMR
Resynthesis Mix
17
HP8 or HP12
Index 1 (i7) Sequencing Primer Mix
18
HP3
Denaturation Solution
19
HT2
Wash Buffer
Table 5 Dual-Indexed Run on a Paired-End Flow Cell
5
Place 15 ml conical tubes filled with 10 ml laboratory-grade water in unused rack positions.
6
Slide the reagent rack into the reagent compartment. Align the rack with the raised guide on the floor of
the compartment.
7
Lower the sippers into the paired-end reagent tubes as follows.
a
Pull the handle towards you and lower it.
b
Inspect the sippers to make sure that they do not bend as they lower into the tubes.
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
34
HiSeq 2000 System Guide