Run Type
Run Length
Run Cycles
Total Cycles
Dual-Indexed Paired-End
2 x 100
101+8+(7)+8+101
225
2 x 92
93+8+(7)+8+93
209
2 x 75
76+8+(7)+8+76
175
2 x 50
51+8+(7)+8+51
125
2 x 35
36+8+(7)+8+36
95
Dual-Indexed Single-Read
1 x 100
101+8+8
117
1 x 75
76+8+8
92
1 x 50
51+8+8
67
1 x 35
36+8+8
52
Sequencing Primers for Nextera Libraries
The Index 1 sequencing primer (HP8) and Read 2 sequencing primer (HP7) provided in the TruSeq v3 cluster
kits are not compatible with Nextera libraries. When sequencing Nextera libraries, use the Index 1
sequencing primer (HP12) and Read 2 sequencing primer (HP11) provided in the TruSeq Dual Index
Sequencing Primer Box.
Performing a dual-index run on a single-read flow cell requires HP9, which is provided in the SR version of the
kit.
Kit Name
Catalog #
TruSeq Dual Index Sequencing Primer Box, SR
FC-121-1003
TruSeq Dual Index Sequencing Primer Box, PE
PE-121-1003
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
7
HiSeq 2000 System Guide