8
Set aside on ice.
Example Calculation
A 101-cycle dual-indexed paired-end run has a total of 225 cycles:
u
124 cycles (101 + 8 + 7 + 8) for Read 1 and the Index Reads
u
101 cycles for Read 2
Read 1 and Index Reads (124 Cycles)
Read 2 (101 Cycles)
124 cycles / 10 cycles = 12.4
• ICB: 12.4 x 4.57 ml = 56.9 ml
• LFN: 12.4 x 0.6 ml = 7.44 ml
• EDP: 12.4 x 0.11 ml = 1.36 ml
101 cycles / 10 cycles = 10
• ICB: 10 x 4.57 ml = 45.7 ml
• LFN: 10 x 0.6 ml = 6 ml
• EDP: 10 x 0.11 ml = 1.1 ml
Use the same calculation for any dual-indexed run. Base the calculation on the total number of cycles in
Read 1 and the Index Reads plus the total number of cycles in Read 2. For dual-indexed paired-end runs,
include the 7 chemistry-only cycles in the Index 2 Read.
Prepare Indexing Reagents
Indexing reagents are used during the indexing reads of an indexed sequencing run. Prepare the reagents
that are appropriate for the run type and library type.
Run Type
Library Type
Index 1 (i7)
Index 2 (i5)
Single-indexed
All libraries except Nextera
HP8
--
Nextera libraries
HP12*
--
Dual-indexed paired-end
All libraries except Nextera
HP8
RMR
Nextera libraries
HP12*
RMR
Dual-indexed single-read
All libraries except Nextera
HP8
HP9*
Nextera libraries
HP12*
HP9*
* Provided in the TruSeq Dual Index Sequencing Primer Box.
Thaw Indexing Reagents
1
Remove the following reagents from -25°C to -15°C storage:
u
For all indexed runs—HP3, HP12 or HP8, and HT2
u
For dual-indexed runs on a single-read flow cell—HP9
u
For dual-indexed runs on a paired-end flow cell—RMR
2
Thaw reagents in a beaker filled with room temperature deionized water for about 20 minutes.
Prepare HT2
1
Invert to mix.
2
Centrifuge at 1000 rpm for 1 minute.
3
Set aside at room temperature.
Prepare HP3 for Indexed Runs
1
Invert to mix, and then pulse centrifuge.