5
Select solution 2 from the drop-down list.
6
Confirm the following default values:
u
Volume: 250
u
Aspirate Rate: 250
u
Dispense Rate: 2000
7
Select Pump.
8
Inspect the flow cell for bubbles passing through the lanes and leaks near the manifolds.
9
If excessive bubbles are present:
a
Check the manifold gaskets for obstructions.
b
Reduce the aspirate rate to 100.
c
Pump another 250 µl of water to the flow cell.
BaseSpace is Unavailable
If BaseSpace is not available, open Windows Services to confirm that the BaseSpace Broker has started. If it
has not started, restart it. If services are running and BaseSpace is still unavailable, contact Illumina Technical
Support.
Stop and Resume a Run
Stopping a run might be necessary if the run was set up incorrectly, if the data quality is bad, or for a
hardware error. When resuming a stopped run, select the appropriate normal stop option that allows the run
to resume.
Stop Option
Real-Time Analysis (RTA)
Option
Able to Resume?
Normal Stop
(End of
Lane\Chemistry)
Keep As Is
Yes. The run resumes at the next chemistry or imaging
command.
Complete For Run
No. The run cannot be resumed.
Complete For Read
Yes. The run resumes at the beginning of the next read.
Normal Stop
(End of Cycle)
Keep As Is
Yes. The run resumes at the next cycle.
Complete For Run
No. The run cannot be resumed.
Complete For Read
Yes. The run resumes at the beginning of the next read.
Immediate Stop
No option
No.
Stop a Run
1
From the Run Overview screen, select Stop to open the Stop Menu.
Figure 18
Stop Menu
2
Select from the following stop options:
u
Normal Stop (End of Lane\Chemistry)—Stops the run only after the current chemistry or imaging
command is complete, and then places the flow cell in a safe state.
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
50
HiSeq 2000 System Guide