Performing Maintenance
FAM 25/45/60/75/95
en(us)
Page 81/160
BEWA FAM25-95 3402054f en(us) 2013-06-13
2013-06-13
See page
24 hours /
daily
500 hours /
monthly
3000 hours /
every six months
6000 hours /
annually
Cleaning the condensation cooler
X
Back-flush water filter combination
(only for version with automatic water
intake)
X
Rotary vane vacuum pump
Checking the oil level
X
Replace oil, oil filter and air de-oiling
element
X
Check and clean the ventilator hood
X
Emptying/Filling Pump
Check and clean the ventilator hood
-
X
Fluid filter
Changing the filter element
X
Emptying and cleaning the filter
housing
-
X
Oil mist separator
Changing the filter element
X
Electric System
Check cable conduits for damage,
replacing as necessary
-
X
Measure the engine power
consumption of all engines and
compare the values with the type label
-
X
Checking the AquaSensor
X
Perform electrical safety inspection in
accordance with DIN VDE 0702 and/or
corresponding national regulation.
-
X
Measure power consumption of the
-
X