GSC53N - GSC57 - ZG47
EN - 59
6.5.3. "2-W", "3-W" and " " faulty cases
If the instrument detect
the
External
Power
supply
adapter
connected to instrument
will show the message
displayed to side.
EARTH
05.06.01
- - -
Vd= ---V
Test:04
RAVG=0.74
REMOVE POWER
3-W
Disconnect the External
Power Supply Adapter
FUNC CLR
If the Instrument detect
a voltage values higher
than 5V the instrument
will shows the screen
displayed alongside.
EARTH
05.06.01
Warning symbol: Voltage
noise in input.
- - -
Vd= 230V
Test:04
RAVG=0.74
VOLT IN INPUT
3-W
FUNC CLR
The
message
"
Rc
high
" indicates that the
instrument can't flow
the minimum current
necessary
for
measurement. Check
that the terminals are
correctly
connected
and the Auxiliary earth
rod connected to B4
(blue conductor) has
not been inserted in a
pebbly
or
poor
conductive ground. If
necessary pour some
water around the rod.
EARTH
05.06.01
- - -
Vd= 1V
Test:04
RAVG=0.74
Rc HIGH
3-W
Message
"Rc":
Check
Auxiliary Earth rods.
FUNC CLR