GSC53N - GSC57 - ZG47
EN - 147
18. APPENDIX 1
–
MESSAGES DISPLAYED
MESSAGE
DESCRIPTION
ADVICES
AUTONOM:
Available memory autonomy for the recording which is
being effected
CLEAR ALL? (Enter)
The operator is trying to cancel all the recordings effected
Press ESC in order not to cancel the whole
memory, press ENTER to confirm
CLEAR LAST? (Enter)
The operator is trying to cancel the last recording effected
Press ESC in order not to cancel the last
recording, press ENTER to confirm
Data saved
The data have been saved
DATA SIZE:
Dimensions of the stored data
HOLD
By pressing the proper key, the HOLD function has been
activated
Press HOLD again to disable this function
Password:
A recording has been started and at least 5 minutes have
passed from the last activity of the instrument (see
paragraph 7).
Insert the password:
F1, F4, F3, F2
Invalid date
The inserted date is not correct
Check the inserted date
Energy Measuring
The instrument is taking an energy measurement
Press F1 to stop it
Memory Full
The memory of the instrument is full
Cancel some recordings after transferring
them to a PC
No ext supply!
A recording has been started without connecting the
external power supply (code A0050; optional for GSC57)
Verify it you really want to start the recording
without the external power supply. In that
case press START again.
No parameter sel
A recording has been started without selecting any value to
be recorded
Press START/STOP and select at least a
value entering the MENU
No Phase selected
Voltage and/or current harmonics have been selected and
the corresponding flag has been enabled (HARMONICS
ON) but no phase voltage or current has been selected
Select at least one phase voltage and/or
current
PASSWORD ERROR
The inserted password is wrong (see paragraph 7).
Check the password
PASSWORD OK
The inserted password is correct
Please wait
The instrument is waiting for the recording to be started
(see paragraph 6)
Recording
The instrument is recording (see paragraph 6)
Too many param
More than 63 parameters have been selected (harmonics
included) or More than 38 parameters with CO-
GENERATION Flag enabled
Deselect some values
Too many records
The quantity of recorded data + Smp exceeds the
maximum allowed (35)
Cancel some recordings after transferring
them to a PC
No Unit selected
ERR: SEQ
The Phase Sequence isn’t correct.
Check the Phase Sequence connection.
ERR: P-
The active powers shown on the right side of the message
are negative
If there isn’t a situation of co-generation
check if the clamps are properly connected
ERR: SEQ & P-
The active powers shown on the right side of the message
are negative and the Phase Sequence isn’t correct.
If there isn’t a situation of co-generation
check if the clamps are properly connected /
check the Phase Sequence connection.
ERR: CONNECTION
The instrument has detected a wrong connection to
Voltage inputs
Check the Voltage connections
Error Vref
The user set a Voltage reference not compatible with
voltage at instrument’s input.
Check Voltage Reference set in “CONFIG
RECORDER”
ERR: SYNC
The System Frequency is out of range
Check the System Frequency, check setting
in ANALYSER CONFIG.
Selection Error
There is a mismatch between the Parameter enabled and
the parameter selected for an AUX recording.
Check the parameter enabled in AUX
position and the selected parameter for
recording.
Error1 Error 5
The instrument memory is damaged.
Contact HT Italia assistance