8.4.4. "WAVE" mode
Selecting the WAVE mode one of the below screens will be displayed according to the
settings made as per paragraph 8.1. The screens show the waveform of the phase or
phase-to-phase voltage.
Example of screen in single-
phase system
Example of screen in “3 wires”
three-phase system
Example of screen in “4 wires”
three-phase system
The symbols used are described in Table 7
For any message displayed see Appendix 1 – MESSAGES DISPLAYED.
Following keys are enabled:
F1
:
(only for three-phase mode): to display the values corresponding to
the following phase.
ESC
:
to return back to METER mode (see paragraph 8.4.2).
SAVE
:
to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and
current present on the instrument inputs. This function is disabled
during a recording.
ENTER/HOLD
: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available.
When the HOLD function is enabled, the word
HOLD
is displayed.
When this function is enabled it’s not possible to record or take an
energy measurement. This function is disabled during a recording or
an energy measurement.
MENU
:
to enter in the
MENU
mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration
MENU during a recording or an energy measurement.
START/STOP
: to record
selected parameters
according to the instrument’s settings
(see chapter 10).
27.09.00 17:35:12
V1 = 230.2 V
Vpk1 = 325.5 V
freq = 50.0 Hz
V
=
27.09.00 17:35:12
5:12
V12 = 400.2 V
Vpk12= 565.5 V
freq = 50.0 Hz
ChgP
27.09.00 17:35:12
V1 = 230.2 V
Vpk1 = 325.5 V
freq = 50.0 Hz
ChgP