GSC53N - GSC57 - ZG47
EN - 16
TEST LEADS
Before each measurement always make sure that
the calibration is referred to the cables in use.
During a continuity test, if the resistance value free
of calibration (that is the resistance value less the
calibration offset value) is
negative
, the symbols
is displayed. Probably the calibration resistance
value stored in the instrument memory is not
referred to the cable in use, therefore a new
calibration must be effected.
6.1.1.1.
Procedure to reset test leads’ calibration parameters
To cancel calibration
parameters
it
is
necessary to effect a
calibration procedure
with a resistance of
test leads higher than
5
(for example with
open test leads). When
a cancellation is effected
the screen nearby
is
displayed first
,
followed
by the screen below:
LOW
05.06.01
Message
>99.9
:
means
that
the
instrument detected a
resistance
higher
than 5 therefore it
will
proceed
with
Reset procedure.
>
99.9
R+ R-
---- ----
---mA ---mA
AUTO 0.11
FUNC CAL