GSC53N - GSC57 - ZG47
EN - 50
6.4.7. LOOP
Faulty Cases
If the instrument detect
the
External
Power
supply
adapter
connected to instrument
will show the message
displayed to side.
LOOP
05.06.01
- - -
---A
FRQ=50.0HZ
VP-N=231V VP-PE=230V
REMOVE POWER
P-N
Disconnect the External
Power Supply Adapter
FUNC Z2
Should the instrument
detect that the phase
and/or neutral cables
are not connected to an
installation,
screen
alongside is displayed
when pressing
START
.
LOOP
05.06.01
- - -
---A
FRQ=50.0HZ
VP-N= 1V VP-PE= 0V
NO VOLTAGE
P-PE 50V
NO VOLTAGE Detected
FUNC Z2
UL
Should the instrument
detect
a
voltage
between
phase
and
neutral higher of 250V,
for example in case the
blue cable is connected
to an installation phase
conductor of a 400V
three-phase system, the
screen
alongside
is
displayed.
LOOP
05.06.01
- - -
---A
Freq=50.0HZ
VP-N=401V VP-PE=230V
HIGH VOLTAGE
P-PE 50V
HIGH VOLTAGE Detected
FUNC Z2
UL