280
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcproc.fb
Chapter 9, 264 Using the Analog Call Processing Subsystem
Using Remote (HP-IB) Control
Example
OUTPUT 714;"RFG:AMPL:STATE OFF"
OUTPUT 714;"DISP RFAN;:RFAN:PME:ZERO"
OUTPUT 714;"RFG:AMPL:STATE ON"
NOTE:
Ensure that no RF power is applied to the
RF IN/OUT
port when the power meter is being
zeroed.
•
Couple the variable frequency notch filter to AFGen1.
This step is only required if audio testing is to be done on the mobile station. This step
couples the variable frequency notch filter to the output frequency of AFGen1 (audio
frequency generator #1). The notch filter is used when making SINAD measurements.
AFGen1 is used to generate the audio tone for the SINAD measurement. Coupling the
notch filter to the audio source ensures the most accurate measurement.
Commands:
OUTPUT 714;"DISP CONF;:CONF:NOTC ’AFGEN1’"
Analog Call Processing Subsystem HP-IB Error Messages
The Analog Call Processing Subsystem HP-IB error messages are numbered 1300
through 1317. See the
“Error Messages”
chapter in the HP 8924C User’s Guide
for details.
Reading An Analog Call Processing Subsystem HP-IB Error Messages
If an error occurs while in the Analog Call Processing Subsystem, an appropriate
error message will be placed in the Error Message Queue. The control program
can read the Error Message Queue to retrieve the error message.
See
“Error Message Queue”
in the Status Reporting chapter of the HP 8924C
User’s Guide for detailed information.
If an error occurred while attempting to decode data messages received from the
mobile station on the reverse control channel or reverse voice channel, the raw
data message bits are displayed in hexadecimal format in the upper right hand
portion of the CALL CONTROL screen.
shows layout of the CALL CONTROL screen when a
decoding error has occurred. The raw data bits can be read by the control program.
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...