28
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcal.fb
Chapter 1, Calibrating the Test Set
Calibrating CDMA Channel Levels
2. Run the PCB_CAL Procedure.
HP-IB Syntax
"TEST:PROC:RUN
"
Manual Operation:
1. Position the cursor next to the Run Test field.
2.Press the knob.
3. Follow instructions on the display. (You will be
instructed to remove all front-panel cables).
4. When the PCB_CAL procedure has completed,
cycle power.
At the beginning of the procedure, the Test Set will beep and the message "Direct latch write occurred. Cycle
power when done servicing" will appear. This is normal.
The PCB_CAL procedure will run for about 8 minutes. During this time the display will show cal factors for I
and Q channels on the screen.When the calibration procedure has completed, the message "Cycle instrument
power to restore test set to normal operating conditions" will be displayed at the top of the screen. At this point
you should cycle power.
TESTS (Main Menu)
ROM
Select Procedure Location:
PCB_CAL
Select Procedure Filename:
RUN TEST
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...