178
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amtxtst.fb
Chapter 4, CDMA Transmitter Tests
Measuring the Range of Open Loop Output Power
3. Select Open Loop power control.
HP-IB Syntax
"DISP CTXR" !displays the CDMA TRANSMITTER CLOSED LOOP RANGE TEST screen.
"CDMA:PCON:MODE ’Open Loop’" !selects Closed Loop power control mode.
Manual Operation:
1. Press and release the SHIFT key, then press the
CDMA SCRNS - RX TEST key to display the
CDMA TRANSMITTER CLOSED LOOP
RANGE TEST screen.
2. Position the cursor at the Closed Loop Pwr Cntl
field and press the knob to display the Choices
menu.
3. Select Open Loop from the list of choices.
When Open Loop is selected, the Test Set sends alternating power control bits on the Forward Traffic
Channel.
CDMA TRANSMITTER CLOSED LOOP RANGE TEST
Open Loop
Closed Loop
Pwr Cntl
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...