150
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amtxtst.fb
Chapter 4, CDMA Transmitter Tests
Measuring Minimum/Maximum Power
Measuring Minimum/Maximum Power
The Test Set provides a quick way to determine an MSUT’s minimum power and
maximum power using both open loop and closed loop power control. Two power
measurements will be displayed at the end of this procedure, which takes several
seconds to complete.
Test Prerequisites
•
The Channel Power measurement should be calibrated at least as often as the following
conditions arise:
•
5 °C change in operating temperature
•
Power cycle
•
Daily
See
"Calibrating Channel Power Measurements" on page 29
if necessary, and
then return to this procedure. If the Channel Power measurement has been calibrated,
zeroing the Average Power measurement is not necessary.
NOTE:
The following procedure assumes that an instrument PRESET will be performed as part of
setting up a call. This clears all settings from other tests that may affect the accuracy of this
test, and allows the procedure to begin from a known instrument state.
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...