Chapter 9, 264 Using the Analog Call Processing Subsystem
Using the CALL CONTROL Screen to Test Call Processing Functions
3. Select the
Release
field. The Test Set will signal the mobile station with a release or-
der. The mobile station will respond to the release order and cease transmission. The
Test Set will terminate transmission on the forward voice channel, and the
Connect
annunciator will turn off. The Test Set will then reconfigure itself for transmission on
the forward control channel, begin to transmit system parameter overhead messages,
and the
Active
annunciator will light.
Change the Transmit Power Level of a Mobile Station
1. Ensure that the Test Set is in the connected state.
See "Page a Mobile Station" on
for information on how to bring a mobile station up on a voice channel.
2. If the CALL CONTROL screen is not displayed, select
CALL CNTL
from the
To
Screen
menu. The CALL CONTROL screen will be displayed.
3. Position the cursor on the
Display
field and select
Meas
. The upper right-hand por-
tion of the CALL CONTROL screen will display modulation quality measurements of
the mobile station’s RF carrier. The Test Set’s data functions, such as the average
(AVG) function, can be used with any of the measurements. Measurement units can
also be changed as desired.
4. Select the
Order
field and select it. A
Choices:
menu is displayed showing the var-
ious power levels which the mobile station can be set to. Select the desired power level
from the
Choices:
menu. The Test Set will then signal the mobile station with an or-
der to set its power level. If the mobile responds properly, the measured value in the
TX
Power
field will change.
5. Position the cursor on the
Display
field and select
Data
.
6. Select the
Order
field. Select the desired power level from the
Choices:
menu. The
Access
annunciator will light while the Test Set signals the mobile station with an or-
der to set its power level. If the mobile responds properly, the message
REVC Return
will flash momentarily in the upper right-hand corner of the CALL CONTROL screen
and then the decoded order confirmation message received from the mobile station will
be displayed. The left-hand
Pwr Lvl:
field will be updated with the new power level
information. The
Connect
annunciator will light when signaling is complete.
Originate a Call from a Mobile Station
1. Ensure that the Test Set is in the active state.
See "Turn On The Test Set’s Control
for information on how to activate the Test Set.
2. Connect the mobile station to the Test Set as shown in
3.
Figure 2, "Connecting a Mobile Station to the Test Set," on page 269
4. Turn on the mobile station.
5. If the CALL CONTROL screen is not displayed, select
CALL CNTL
from the
To
Screen
menu. The CALL CONTROL screen will be displayed.
6. Position the cursor on the
Display
field and select
Data
.
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...