41
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcal.fb
Chapter 1, Calibrating the Test Set
Correcting for RF Path Loss
Cha
p
te
r 1
Calib
ra
ting
The
Tes
t Set
Correcting for RF Path Loss
Approximate time: N/A (this procedure is simply a field entry).
The settings you make in the following procedure must be re-entered after a
power-cycle, instrument preset, or HP-IB reset ("*RST).
It is highly recommended that RF path loss is corrected for in the following
manner.
NOTE:
The Test Set’s attenuator auto-ranging algorithm, used for adjusting gain in the RF analyzer
path, estimates the expected power level from the phone using the open loop power control
formula. External path loss, entered in the procedure below, is used by the auto-ranging
algorithm to ensure the analyzer is not overdriven or underdriven.
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...