29
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcal.fb
Chapter 1, Calibrating the Test Set
Calibrating Channel Power Measurements
Cha
p
te
r 1
Calib
ra
ting
The
Tes
t Set
Calibrating Channel Power Measurements
Approximate time: 2 minutes
Channel power measurements should be calibrated whenever any of the
following events occur:
•
After a 30-minute warm-up period
•
After firmware is upgraded
•
When the "Uncal" annunciator is flashing
•
If the RF connections to the PCS Interface are adjusted
•
If the ambient temperature drifts more than 5 deg C after a 30-minute warm-up period
Channel Power measurements will be calibrated for the combined frequency
bands included in the RF Chan Std (RF Channel Standard) and Alt Chn Std
(Alternate Channel Standard) field selections.
Average Power measurements are zeroed as part of the Channel Power calibration
process.
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...