303
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcproc.fb
Chapter 9, 264 Using the Analog Call Processing Subsystem
Using the CALL DATA Screen
Cha
p
te
r 9
Usin
g the
Anal
og Ca
ll
Proce
s
s
ing Su
bsy
s
tem
The page does not have to happen while you are viewing the CALL DATA
screen. When the CALL DATA screen is viewed, the contents of the display
words will be the decoded data from the last successful call processing function
(registration, origination, page, release, order).
Viewing the Decoded
Reverse Channel
Words from a Page to a
Mobile Station That
Has Not Registered
with the Test Set
1. If not already done, condition the Test Set.
See "Conditioning the Test Set for Call
for further information.
2. If not already done, configure the Test Set
See "Configure the Test Set" on page 286
3. Connect the mobile station to the Test Set as shown in
bile Station to the Test Set," on page 269
.
4. Turn on the mobile station.
5. If the CALL CONTROL screen is not displayed, select
CALL CNTL
under the
To
Screen
menu. The CALL CONTROL screen will be displayed.
6. Select the lower
MS Id
field. Enter the mobile station identification number, either the
phone number or the MIN number depending upon what the upper
MS Id
field is set to.
7. Select
CALL DATA
from the
To Screen
menu. The CALL DATA screen will be
displayed.
8. Select the
Active
field. The
Active
annunciator will light when the control channel
is turned on.
9. Select the
Page
field. The
Page
annunciator will light while the Test Set pages the
mobile on the forward control channel. If the mobile station responds properly on the
reverse control channel, the message
RECC Return
will flash momentarily in the up-
per right-hand corner of the CALL CONTROL screen. The
Access
annunciator will
then light while the Test Set sends the mobile station an alert order on the assigned
voice channel. The mobile station should ring.
10. Press the SEND key on the mobile station. The
Connect
annunciator will light if the
mobile station properly signals the Test Set when the SEND key is pressed. The mobile
station is now connected to the Test Set on the assigned voice channel.
11. Select the
Display Word
field. A list of reverse channel words which can be dis-
played will appear in a
Choices:
menu. Select the desired word from the list of
choices. The decoded contents for each of the fields in the selected word will be dis-
played in the lower portion of the display.
NOTE:
If a word is chosen which was not part of the decoded reverse channel’s message stream,
all the fields will be blank when it is displayed.
Step 9 and step 11 in the above procedure can be reversed, that is - the desired
word can be selected first, then the mobile station can be paged. If the page is
successful, the contents of each display word contained in the reverse control
channel’s message stream will be updated when the decoding is complete.
The page does not have to happen while you are viewing the CALL DATA
screen. When the CALL DATA screen is viewed, the contents of the display
words will be the decoded data from the last successful call processing function
(registration, origination, page, release, order).
Содержание HP 8924C
Страница 21: ...Contents 21 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC Table of Contents Index 351 ...
Страница 22: ...Contents 22 N mkt MANUALS HP8924C APPMOD BOOK APPMOD TOC ...
Страница 51: ...51 N mkt MANUALS HP8924C APPMOD BOOK chapters amcall fb Chapter 2 Setting Up a Call 2 Setting Up a Call ...
Страница 71: ...71 N mkt MANUALS HP8924C APPMOD BOOK chapters amrcvrt fb Chapter 3 CDMA Receiver Tests 3 CDMA Receiver Tests ...
Страница 139: ...139 N mkt MANUALS HP8924C APPMOD BOOK chapters amtxtst fb Chapter 4 CDMA Transmitter Tests 4 CDMA Transmitter Tests ...
Страница 199: ...199 N mkt MANUALS HP8924C APPMOD BOOK chapters amhoff fb Chapter 5 CDMA to Analog Handoff 5 CDMA to Analog Handoff ...
Страница 317: ...317 N mkt MANUALS HP8924C APPMOD BOOK chapters amsrq fb Chapter 10 Controlling Program Flow 10 Controlling Program Flow ...