Programming the Enhanced Monitor
101
Chapter 3
STATus:QUEStionable:CURRent:CONDition?
The
STATus:QUEStionable:CURRent:CONDition?
query returns an integer
representing the contents of the condition register associated with the CURRent
status group.
The condition register continuously monitors the hardware and firmware status of
the instrument. There is no latching or buffering for this register; it is updated in real
time. Condition registers are read-only.
The following table describes each bit in the CURRent status group:
Returned Data
Comments
•
A bit is set when a current measurement is outside of the user-set limit.
Error conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
*RST does not affect the status system.
Related
Commands
STATus:QUEStionable:CURRent:EVENt?,
STATus:QUEStionable:CURRent:ENABle
Bit
Description
Decimal Value
0
+24 (P24) volt power supply
1
1
+12 (P12) volt power supply
2
2
+5 (P5) volt power supply
4
3
Unused
8
4
Unused
16
5
-2 (N2) volt power supply
32
6
-5.2 (N5P2) volt power supply
64
7
-12 (N12) volt power supply
128
8
-24 (N24) volt power supply
256
9
Unused 512
10
Unused
1024
11
Unused
2048
12
Unused
4096
13
Unused
8192
14
Unused
16384
15
Unused
32768
Type
Range
Default
Description
int16
0 - 32767 none
Bit mask indicating which conditions are set (1) or cleared (0)
for the CURRent status group.
Number
Message
Probable Cause
-241
“Hardware missing”
Power supply doesn’t support current measurements
Содержание E8402A
Страница 8: ...10 Contents ...
Страница 12: ...14 Notes ...
Страница 14: ......
Страница 26: ...28 Getting Started Chapter 1 ...
Страница 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Страница 214: ...216 Servicing Your Mainframe Chapter 5 ...
Страница 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...