234
Rack Mounting and Option Installation
Appendix B
5. Attach the rack mount adapters to the mainframe using the m5x8 pan
head screws provided. To mount the mainframe flush with the rack
front, begin with the 3rd hole from the front of the adapter (Figure
B-4).
With the standard adapters (Option 923), the mainframe can be recess
mounted up to 270.7 mm (10.6 inches), or extended out from the rack
up to 147.6 mm (5.8 inches) in 12.3 mm (approximately 1/2 inch)
increments.
The rack mount adapters for the VXI compliant and flush mount
options are also shown in Figure B-4.
Figure B-4. Attaching the Rack Mount Adapter to the Mainframe.
m5x8 Pan Head
Screws
Recess Mount in 12.3
(~1/2 inch) Increments
Standard
Adapter
Option 923
Adapter
Compliant
VXI Plug & Play
Option 924
Adapter
Flush Mount
Option 925