Programming the Enhanced Monitor
109
Chapter 3
STATus:QUEStionable:POWer:LIMit
<limit>
STATus:QUEStionable:POWer:LIMit?
[MIN|MAX]
The
STATus:QUEStionable:POWert:LIMit
sets the limit for wattage for the total of
the seven supplies in the mainframe. If it is exceeded, a warning will be issued. This
value is stored in non-volatile memory.
The
STATus:QUEStionable:POWert:LIMit?
returns a floating point number
representing the limit for total power in the mainframe.
Parameters
Returned Data
Comments
•
Remember to execute a
SYSTem:NVSave
after setting all the limits you
desire. This will move them into non-volatile memory.
•
This command determines if <value> is within the valid range for the specified
<supply>; if it is not, the limit is set to MAX. No error is generated. Use MIN
and MAX parameters with the query to determine the maximum and minimum
power limits.
Reset Condition
*RST resets register the last value saved by the SYST:NVS command.
Related
Commands
STAT:QUES:CURR:LIMit?, SYSTem:NVSave
Name
Type
Range
Default
Description
<
limit>
int16
0 to 2000,
MIN, MAX
none
Total power limit. Different power
supplies have different maximums.
optional
parameter
enum
MIN, MAX
Returns the minimum or maximum power
limit.
Type
Range
Description
float
0 to 2000
Total power limit.
Содержание E8402A
Страница 8: ...10 Contents ...
Страница 12: ...14 Notes ...
Страница 14: ......
Страница 26: ...28 Getting Started Chapter 1 ...
Страница 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Страница 214: ...216 Servicing Your Mainframe Chapter 5 ...
Страница 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...