Programming the Enhanced Monitor
Chapter 3
HISTory:CURRent[:HISTogram]?
<
supply
>[,MIN|MAX]
HISTory:CURRent[:HISTogram]?
<supply> returns the histogram data held for the
specified <supply>. Ten values are returned for the amount of time the <supply>’s
current spent in ten different current ranges. Those current ranges can be queried
with the MIN and MAX optional parameter. The units of the time values returned
are HOURs by default, but can be changed with the
HISTory:UNIT[:TIME]
command.
The units returned for current ranges queried with the MIN or MAX parameters are
integer values in milliamps.
Parameters
Returned Data
Comments
•
HIST:CURR? P5
returns 10 time values for the +5V supply.
•
HIST:CURR? P5,MAX
returns 10 values representing the maximum currents in
each current range in milliamps.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
*RST has no effect on the HISTory subsystem.
Related
Commands
HISTory:CURRent:MAXimum?, HISTory:RESet:CURRent,
HISTory:UNITs[:TIME]
Name
Type
Range
Default
Description
<
supply>
enum
P5, P12, N12, P24,
N24, N5PT2, N2
none
Selects the power supply whose current
histogram is desired. P5 is +5 Vdc
supply, N12 is -12 Vdc supply, etc.
optional
parameter
enum
MIN, MAX
none
Returns the minimums or maximums of
the current ranges in milliamps.
Type
Range
Default
Description
<range1>
int32
0-2147483647
none
lowest current range.
<range2>
int32
0-2147483647
none
second current range.
<range3>
int32
0-2147483647
none
third current range.
<range4>
int32
0-2147483647
none
fourth current range.
<range5>
int32
0-2147483647
none
fifth current range.
<range6>
int32
0-2147483647
none
sixth current range.
<range7>
int32
0-2147483647
none
seventh current range.
<range8>
int32
0-2147483647
none
eighth current range.
<range9>
int32
0-2147483647
none
ninth current range.
<range10>
int32
0-2147483647
none
highest current range.
Number
Message
Probable Cause
-109
“Missing Parameter”
The
<supply> parameter was omitted.
-224
“Illegal Parameter”
The <supply> or optional parameter not correct.
-221
“Settings Conflict”
Power supply doesn’t support current measurements.
-311
“Memory error”
The historical data’s memory is not available.
Содержание E8402A
Страница 8: ...10 Contents ...
Страница 12: ...14 Notes ...
Страница 14: ......
Страница 26: ...28 Getting Started Chapter 1 ...
Страница 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Страница 214: ...216 Servicing Your Mainframe Chapter 5 ...
Страница 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...