Programming the Enhanced Monitor
97
Chapter 3
STATus:QUEStionable:BLOWer:EVENt?
The
STATus:QUEStionable:BLOWer:EVENt?
query returns an integer representing
the value of the event register for the BLOWer status group.
The event register latches positive transition events from the blower condition
register. A positive transition occurs when a condition makes a transition from a low
to a high state. Bits in the event register are latched, and once set, they remain set
until cleared by a
STATus:QUEStionable:BLOWer:EVENt?
query or *CLS (clear
status). There is no buffering; so while an event bit is set, subsequent events
corresponding to that bit are ignored. Event registers are read only.
Returned Data
Comments
•
Refer to
STATus:QUEStionable:BLOWer:CONDition?
for a description of each
bit in the BLOWer status group.
•
Cleared by: *CLS, power-on, and by reading the event register.
Reset Condition
*RST does not affect the status system, use *CLS to clear all event registers.
Related
Commands
STATus:QUEStionable:BLOWer:CONDition?,
STATus:QUEStionable:BLOWer:ENABle, *CLS
Type
Range
Default
Description
int16
0 - 32767
none
Bit mask indicating which events are set (1)
or cleared (0) for the BLOWer status group.