Programming the Enhanced Monitor
91
Chapter 3
STATus:OPERation:CONDition?
The
STATus:OPERation:CONDition?
query returns an integer representing the
contents of the condition register associated with the operation status group. The
condition register continuously monitors the hardware and firmware status of the
instrument. There is no latching or buffering for this register; it is updated in real
time. Condition registers are read-only. The following table describes each bit in the
OPERation status group:
Returned Data
Comments
•
The condition register reflects the real-time state. This differs from events
which are latching.
Reset Condition
*RST does not affect the status system.
Related
Commands
STATus:OPERation:EVENt?, STATus:OPERation:ENABle
Bit
Description
Decimal
Value
0
CALibrating. While a “1”, the instrument is calibrating.
1
1
Unused
2
2
Unused
4
3
Unused
8
4
Measuring. While a “1”, the instrument is measuring.
16
5
Unused
32
6
Unused
64
7
Unused
128
8
Unused
256
9
Mainframe power-down. While a “1”, the instrument is in the
power-down state (command execution is only available through the
Enhanced Monitor’s RS-232 port with ex5V supply).
512
10
History Queue full. While a “1”, history queue is full and is accepting no
more events.
1024
11
Front Panel Keys. Momentarily set to “1” when monitor detects any front
panel key pressed. Use Event Register to monitor events.
2048
12
Unused
4096
13
Reserved 8192
14
Reserved 16384
15
Always Zero
32768
Type
Range
Default
Description
int16
0 - 32767 none
Bit mask indicating which conditions are set (1) or cleared (0)
in the OPERation status group. Each condition has a decimal
weighted value.
Содержание E8402A
Страница 8: ...10 Contents ...
Страница 12: ...14 Notes ...
Страница 14: ......
Страница 26: ...28 Getting Started Chapter 1 ...
Страница 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Страница 214: ...216 Servicing Your Mainframe Chapter 5 ...
Страница 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...