150
Programming the Enhanced Monitor
Chapter 3
SYSTem:NVSave
SYSTem:NVSave causes the enhanced monitor to store many of its parameters to
non-volatile memory. This includes temperature limits, current limits, total power
limits, status enable settings, PSC flag, beep frequency, beep duration, system name,
mainframe serial number, and all RS-232 settings.
Comments
•
Non-volatile memory has a finite access lifetime. For this reason, these values
are saved only when explicitly requested with this command and all at one
time. The commands/registers affected are:
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using SYSTem:ERRor? query.
Reset Condition
*RST has no effect
Related
Commands
SYSTem:NVRecall
Register
Register
STATus:OPERation:ENABle
SYST:BEEP:FREQ
STATus:QUEStionable:ENABle
SYST:BEEP:TIM
STAT:QUES:BLOWer:ENABle
SYST:NAME
STAT:QUES:CURRent:ENABle
SYST:SNUM
STAT:QUES:TEMPerature:ENABle
HIST:UNIT:TIME
STAT:QUES:VOLT:ENABle
*PSC
STAT:QUES:VOLT:PTR
*ESE
STAT:QUES:CURR:LIM
*SRE
STAT:QUES:TEMP:LIM
SYST:COMM:SER:REC:BAUD
STAT:QUES:POW:LIM
SYST:COMM:SER:REC:BITS
SYST:COMM:SER:CONT:RTS
SYST:COMM:SER:REC:PACE
SYST:COMM:SER:ECHO
SYST:COMM:SER:PARity
SYST:COMM:SER:ERES
SYST:COMM:SER:SBIT
SYST:COMM:SER:LBUF
Number
Message
Probable Cause
-221
“Settings Conflict”
The mainframe’s power is off.
-311
“Memory Error”
Non-volatile memory is inaccessible.
Содержание E8402A
Страница 8: ...10 Contents ...
Страница 12: ...14 Notes ...
Страница 14: ......
Страница 26: ...28 Getting Started Chapter 1 ...
Страница 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Страница 214: ...216 Servicing Your Mainframe Chapter 5 ...
Страница 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...